Ninety-two subjects were recruited and 85 subjects were analyzed. For validation criterion 1, the mean ± SD of the differences between the test device and reference BP readings was 0.3 ± 7.2/2.2 ± 5.5 mmHg (systolic/diastolic). For validation criterion 2, the SD of the averaged BP differences between the test device and reference BP per subject was 6.1/4.8 mmHg (systolic/diastolic).
Conclusion: The YuWell YE660D oscillometric upper-arm electronic BP monitor has passed the requirements of the AAMI/ESH/ISO Universal Standard (ISO 81060-2 : 2018) and its Amendment 1. 2020 in adults and hence can be recommended for home and clinical use