X-Ray Generator
Input power
220VAC, single phase
Nominal electric power
32kW
Tube voltage range
40~150kV
Current output range
10~400mA
Time of exposure
1ms~10000ms
Current & Time range
0.1~400mAs
Automatic APR linkage function
Provide
2
X-Ray Tube
Focal spot
0.6/1.2mm
Focal power
20kW/50kW
Tube voltage range
40~150kV
Anode heat capacity
200kHU
Rotary anode
2800 RPM
3
Flat Panel Detector
FPD type
a-Si/Csl
Image area
17×17inch
Acquisition matrix
3072×3072
Pixels size
139 μm
Resolution
3.6lp/mm
3
A / D conversion
16bit
Imaging time
≤9s
Image review time
≤7s
4
Stand Assembly
Moving range of detector under bed
600mm, deviation±5%
Height of tube stand
2310mm, deviation±5%
Tube stand rotation angle
±180°, deviation±2%
Tube stand horizontal movement range
1775mm, deviation±5%
Tube rotation angle
±180°, deviation±2%
Tube vertical movem